from the Blog Material Compatibility in Multi-Layer Wearable Device Design


Wearable medical devices often rely on complex constructions that combine flexible substrates with electronic components and skin-contacting materials. Each layer has a specific function, yet its performance depends on how it interacts with the materials surrounding it.

This can create a materials engineering challenge, for example, a material may perform as expected in isolation, but its properties can change when incorporated into a complete assembly. Chemical interactions or mechanical stress can affect the interfaces between layers, potentially compromising device performance over time.

Material compatibility should therefore be considered from the early stages of product development. Adhesives are a particularly important part of this assessment, as they form the bonding interfaces between materials with different properties.

Understanding the Layer Structure of Wearable Medical Devices

The construction of a wearable medical device varies according to its application. Many designs incorporate several distinct layers, each contributing to the overall performance of the device.

A typical construction may include a skin-contact layer, a medical adhesive and a flexible substrate. An electronic or sensor layer may sit beneath these components, with additional encapsulation materials providing protection.

The skin-contact layer influences comfort and the way the device responds to prolonged exposure to moisture. Where an adhesive forms part of this interface, its performance needs to be assessed alongside the material it contacts.

Substrate or carrier films provide a supporting structure for the device and both their flexibility and surface energy can influence how effectively adjoining layers bond. Differences in mechanical behaviour can also create stress at the interfaces when the device bends or stretches.

Sensor and electronic layers introduce further considerations. These components may be sensitive to moisture ingress, making the surrounding protective materials particularly important and the interaction with underlying materials also requires assessment.

Some designs incorporate conductive or barrier layers to provide additional functionality. Each introduces another interface that may influence the behaviour of the complete assembly.

Material Compatibility Challenges

Chemical Interactions & Material Migration

Some materials contain components that can migrate into adjacent layers over time. Plasticisers and other low-molecular-weight components can migrate between adjacent materials, potentially altering their properties and affecting the performance of adhesive interfaces.

Absorption and diffusion can also influence material behaviour. An adhesive may absorb substances from a neighbouring substrate or coating, changing its properties over the intended service life.

These effects can become more apparent following prolonged contact or environmental exposure, making ageing behaviour an important consideration during material selection.

Moisture & Environmental Exposure

Wearable devices may experience regular exposure to sweat and humidity. Moisture can affect polymer properties and weaken certain bonding interfaces. Ingress through protective layers may also expose sensitive electronic components.

Materials with different coefficients of thermal expansion respond differently to changes in temperature. This can create stress at bonded interfaces, particularly where the device also experiences repeated movement.

Mechanical flexibility is another factor to consider as a wearable may undergo frequent bending or stretching during normal use. Materials with different elastic properties can respond differently to these movements, increasing stress across the assembly.

Surface Characteristics & Coatings

Surface energy influences how effectively an adhesive wets a substrate and establishes contact. Materials with low surface energy can present particular bonding challenges, potentially requiring surface treatment or a suitable adhesive system.

A coating applied to improve a material’s functional properties may change the surface available for bonding too. Its compatibility with the adhesive therefore needs to be assessed as part of the material selection process.

Sterilisation Considerations

Where a wearable device requires sterilisation, the chosen process becomes an additional factor in material selection. Exposure conditions can affect polymer properties or adhesive performance, depending on the materials involved.

The Role of Adhesive Selection

Looking Beyond Bond Strength

Initial bond strength is a measure of adhesive performance, but it provides only part of the picture. Adhesion describes the bond between the adhesive and a substrate, while cohesion refers to the internal strength of the adhesive itself.

Both influence the integrity of the assembled device in the way that an adhesive may achieve a strong initial bond but experience reduced performance following repeated flexing or prolonged exposure to moisture. This could include changes in tack, peel adhesion or cohesive integrity over time.

The adhesive also needs to accommodate the movement of the surrounding materials while maintaining the required bond throughout the intended period of use.

Compatibility With Skin-Contacting Materials

Where an adhesive forms part of the patient-contacting interface, its suitability needs to be evaluated in relation to the intended contact conditions.

Wear duration can influence the requirements, particularly where a device remains attached to the skin for extended periods. Biocompatibility must be assessed in accordance with ISO 10993, considering the finished medical device and the nature, duration and conditions of patient contact.

An adhesive’s biocompatibility alone does not establish the biological safety of the finished device, which must be assessed as a whole.

Manufacturing Requirements

Adhesive selection also affects the manufacturing process as application conditions and curing requirements can influence production consistency. The chosen adhesive needs to suit the intended assembly method and the properties of the materials being joined.

The most suitable adhesive is one that performs effectively within the complete material system. Its selection should take place alongside substrate and coating decisions during product development.

BDK works with medical device manufacturers to assess adhesive and material compatibility, helping develop constructions that meet application requirements and can be manufactured consistently. Early engagement can help identify potential interface challenges while the layer structure is still being developed.

Common Material Compatibility Failure Modes

Compatibility issues can affect the performance of individual layers and, in some cases, compromise the wider assembly.

  • Adhesive failure occurs when the bond separates from the substrate, often following environmental exposure or mechanical stress.
  • Cohesive failure occurs within the adhesive itself and indicates that the internal strength of the adhesive has become insufficient for the applied load.
  • Substrate failure happens where the bonded material tears while the adhesive interface remains intact.
  • Material mismatch can cause curling or distortion.
  • Differences in thermal expansion may create internal stresses that affect the shape of the device.
  • Ageing and environmental exposure can contribute to cracking or embrittlement in certain materials.
  • A loss of flexibility can place additional strain on neighbouring layers during repeated movement which can result from ageing, environmental exposure or material interactions.
  • Changes in adhesive properties may lead to reduced tack.
  • Migration between materials can alter the performance of an interface over time, potentially resulting in unwanted residue.

These failures can have consequences beyond the immediate bonding issue. Understanding these pathways helps manufacturers establish appropriate testing requirements during development.

Testing & Validation Approaches

Compatibility testing should begin during early R&D, when manufacturers have greater flexibility to assess alternative materials and adhesive systems.

Adhesion & Mechanical Testing

Peel and shear tests can help assess how a bonded assembly responds to different loading conditions. The selected method should reflect the way the device is expected to experience force during use.

For flexible devices, repeated bending or stretching can place cumulative stress on bonded interfaces. Cyclic loading is therefore a useful part of the evaluation where appropriate.

Environmental Conditioning & Ageing

Environmental conditioning can reveal changes that may occur during the device lifecycle. Testing may involve temperature and humidity exposure, or sweat and moisture analysis, depending on the intended application.

Comparing results before and after conditioning can help identify changes in material behaviour. Accelerated ageing can also provide further insight into how materials may behave over time.

Biocompatibility Assessment

Biocompatibility assessment should form part of the evaluation for patient-contact materials. This requires consideration of the biological safety profile of the relevant materials, with ISO 10993 providing a recognised framework for evaluation.

Where sterilisation applies, relevant material and adhesive properties should also be assessed before and after processing.

Testing the Complete Assembly

Individual material datasheets provide useful starting information, but they offer limited insight into interactions within the finished construction. Testing representative assemblies allows manufacturers to evaluate how the layers behave together under conditions that reflect the intended application.

Design & Manufacturing Considerations

Material compatibility has practical implications throughout device development and production.

  • Surface Preparation: Cleaning or treating a substrate can influence how effectively an adhesive establishes contact. Consistent preparation is essential when developing a repeatable bonding process.
  • Bonding Parameters: Application conditions, pressure and temperature may influence the resulting interface, depending on the adhesive technology. Layer thickness can also affect the final construction, particularly in devices where a slim profile is required.
  • Assembly Sequence: The order in which layers are joined can influence their exposure to processing conditions, including heat or curing stages.
  • Production Scale-Up: A material combination that performs successfully during laboratory development may require additional process evaluation when transferred to manufacturing equipment. Variations in application conditions can affect consistency between production batches.
  • Material Changes: A revised substrate specification or coating may alter an established bonding interface. Even where the replacement material appears similar, its surface characteristics could influence adhesive performance.
  • Converting Process: Lamination conditions, web tension, material alignment and cutting tolerances can all influence the performance of multi-layer constructions. Understanding these factors during development helps ensure that a design can be translated into consistent, repeatable production.

Developing a Material Compatibility Strategy

A structured approach can help manufacturers identify potential problems early and reduce the risk of late-stage design changes.

The first step is to map the complete material stack which should identify each material within the construction, including coatings and adhesives, alongside the interfaces where direct contact occurs.

Performance requirements can then be established for each interface, reflecting the mechanical loads expected during manufacturing and use, together with relevant environmental conditions.

As development progresses, testing should move towards representative assemblies. This provides a clearer understanding of how the materials interact under realistic conditions.

Material specifications and supplier controls should also be established to support production consistency. Any subsequent design or process changes should prompt a review of the existing compatibility assessment.

Involving adhesive and materials specialists early in this process can help manufacturers evaluate potential issues while the device design is still being developed.

Designing the Whole Material System

Material compatibility is a fundamental consideration in multi-layer wearable device development. The performance of each material depends partly on its interaction with neighbouring layers and the conditions experienced by the finished assembly.

Adhesive selection plays a central role in this process. A considered approach to material selection and validation helps manufacturers reduce the risk of late-stage design changes.

BDK works with medical device manufacturers to develop and convert specialist adhesive constructions, helping customers assess material compatibility, optimise bonding performance and achieve consistent manufacturing results. Contact us today to find out more information and to discuss your next project with us.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.